ECG
Input:
3 / 5 / 10 wires ECG cable
Lead section:
I, II, III, aVR, aVL, aVF, V1-V6
Gain selection:
×1/4, ×1/2, ×1, ×2, ×4 and automatic
Sweep speed:
3.125, 6.25, 12.5, 25, 50mm/s
Heart rate range:
15–350 bpm
Calibration:
±1mV
Accuracy:
±1 bpm or ±1%
SPO2
Display type:
Waveform, Data
Measurement range:
0–100%
Accuracy:
±3% (between 70%–100%)
SpO2 resolution:
1%
Pulse rate range:
25–250 bpm
Accuracy:
±1 bpm or ±2%
PR Resolution:
1 bpm
2-Respiration (Impedance & Nasal Tube)
Measurement range:
0–120 rpm
Accuracy:
±1 rpm or ±5%
Resolution:
1 rpm
Anesthesia Gas Analysis (Option)
Method:
Infrared Absorption
Gas Sorts:
CO2, N2O, Des, Iso, Enf, Sev, Hal and optional O2
Measurement range:
CO2: 0–13.6%; N2O: 0–100%;
O2: 0–100%; Hal: 0–8%;
Iso: 0–8.5%; Enf: 0–10%;
Sev: 0–10%; Des: 0–20%
RR:
0–100 rpm
Data input:
Fi and Et values
NIBP
Philosophy:
Oscillometric method
Measurement parameter:
Systolic, Diastolic, Mean
Measurement method:
Manual, Auto, Continuous
Measuring range:
SYS: 30–270mmHg
DIA: 10–200mmHg
MAP: 20–220mmHg
Resolution:
1mmHg
Accuracy:
Max average error ±5mmHg
2-IBP (Option)
Measurement range:
-100–300mmHg
Number of channels:
2 channel
Accuracy:
±1mmHg or ±2%
IBP Pressure Tag:
ART, CVP, PVP, LAP, RAP, PAP, ICP, LVP, etc.
2-Temperature (Rectal & Surface)
Number of channel:
2 channel
Measurement range:
0°C–60°C
Accuracy:
±0.2°C
Resolution:
0.1°C
Display:
T1, T2, ΔT
ETCO2-Plug & Play (Option)
Measurement type:
Main/Side stream
Measurement range:
0%–19.7%, 0–150mmHg
Accuracy:
0–40mmHg ±2mmHg
41–70mmHg ±5% of reading
71–100mmHg ±8% of reading
101–150mmHg ±10% of reading
Cardiac Output (Option)
Measuring Method:
Thermodilution
Measurement range:
C.O.: 0.1–20.0L/min
TB:
20.0°C–45.0°C
TI:
0.0°C–30.0°C
Accuracy:
C.O. ±5% or ±0.1L/min
Resolution:
C.O.: 0.1L/min