Features

• Alarm Function: Audible and visual alarms with adjustable limits;
• Data: Trend data storage for up to 360 hours;
• Printing: Printing for data, waveform, Trend tables and trend diagrams;
• User Interface: 12.1" color TFT LCD screen, available in 15 languages;
• Versatility: Suitable for adults, pediatrics, and neonates; widely applicable in physician's offices, OR/OT, ICU, CCU, NICU, clinics, etc;
• Communication: WAN communication function to network with Central Monitoring System, supporting both wired and wireless networking and compatibility with HL7 systems;
• Accuracy Assurance: "Auto-current drift balanced system" ensures accuracy, along with unique patents for impedance RESP and ECG waveform stability during electrosurgical procedures;
• Advanced Functions: Includes ST-segment analysis, bed-to-bed view, venipuncture support, pacemaker monitoring, arrhythmia analysis and 5 calculations (Hemodynamic, Drug Dose, Oxygenation, Renal Function, Ventilation);
• Specialized Tests: Offers unique nasal tube respiration testing suitable for patients with burns, upper airway obstructions, and newborns.

Technical Specifications

ECG
Input: 3 / 5 / 10 wires ECG cable
Lead section: I, II, III, aVR, aVL, aVF, V1-V6
Gain selection: ×1/4, ×1/2, ×1, ×2, ×4 and automatic
Sweep speed: 3.125, 6.25, 12.5, 25, 50mm/s
Heart rate range: 15–350 bpm
Calibration: ±1mV
Accuracy: ±1 bpm or ±1%
SPO2
Display type: Waveform, Data
Measurement range: 0–100%
Accuracy: ±3% (between 70%–100%)
SpO2 resolution: 1%
Pulse rate range: 25–250 bpm
Accuracy: ±1 bpm or ±2%
PR Resolution: 1 bpm
2-Respiration (Impedance & Nasal Tube)
Measurement range: 0–120 rpm
Accuracy: ±1 rpm or ±5%
Resolution: 1 rpm
Anesthesia Gas Analysis (Option)
Method: Infrared Absorption
Gas Sorts: CO2, N2O, Des, Iso, Enf, Sev, Hal and optional O2
Measurement range: CO2: 0–13.6%; N2O: 0–100%;
O2: 0–100%; Hal: 0–8%;
Iso: 0–8.5%; Enf: 0–10%;
Sev: 0–10%; Des: 0–20%
RR: 0–100 rpm
Data input: Fi and Et values
NIBP
Philosophy: Oscillometric method
Measurement parameter: Systolic, Diastolic, Mean
Measurement method: Manual, Auto, Continuous
Measuring range: SYS: 30–270mmHg
DIA: 10–200mmHg
MAP: 20–220mmHg
Resolution: 1mmHg
Accuracy: Max average error ±5mmHg
2-IBP (Option)
Measurement range: -100–300mmHg
Number of channels: 2 channel
Accuracy: ±1mmHg or ±2%
IBP Pressure Tag: ART, CVP, PVP, LAP, RAP, PAP, ICP, LVP, etc.
2-Temperature (Rectal & Surface)
Number of channel: 2 channel
Measurement range: 0°C–60°C
Accuracy: ±0.2°C
Resolution: 0.1°C
Display: T1, T2, ΔT
ETCO2-Plug & Play (Option)
Measurement type: Main/Side stream
Measurement range: 0%–19.7%, 0–150mmHg
Accuracy: 0–40mmHg ±2mmHg
41–70mmHg ±5% of reading
71–100mmHg ±8% of reading
101–150mmHg ±10% of reading
Cardiac Output (Option)
Measuring Method: Thermodilution
Measurement range: C.O.: 0.1–20.0L/min
TB: 20.0°C–45.0°C
TI: 0.0°C–30.0°C
Accuracy: C.O. ±5% or ±0.1L/min
Resolution: C.O.: 0.1L/min

Configuration

Standard configuration

ECG, NIBP, SPO2, 2RESP, 2TEMP, HR/PR, Rechargeable battery

Optional configuration

Built-in Thermal Printer, 12-Lead ECG, Nellcor SPO2, Masimo SPO2, 2-IBP, Suntech NIBP, ETCO2, CO, Anesthesia Gas Analysis, Touch Screen, Wireless Network, 12-Volt Power Supply, High Capacity Battery, Carry Bag, Wall Mount, Moving Cart.

[* If you desire multiple configurations, please contact us to confirm accessibility.]